Приказ основних података о документу
Optimizing moving average control procedures for small-volume laboratories: can it be done?
dc.creator | Lukić, Vera | |
dc.creator | Ignjatović, Svetlana | |
dc.date.accessioned | 2019-11-15T11:48:19Z | |
dc.date.available | 2019-11-15T11:48:19Z | |
dc.date.issued | 2019 | |
dc.identifier.issn | 1330-0962 | |
dc.identifier.uri | https://farfar.pharmacy.bg.ac.rs/handle/123456789/3467 | |
dc.description.abstract | Introduction: Moving average (MA) means calculating the average value from a set of patient results and further using that value for analytical quality control purposes. The aim of this study was to examine whether the selection, optimization and validation of MA procedures can be performed using the already described bias detection simulation method and whether it is possible to select appropriate MA procedures for a laboratory with a small daily testing volume. | en |
dc.language.iso | en | sr |
dc.publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine | sr |
dc.relation | info:eu-repo/grantAgreement/MESTD/Basic Research (BR or ON)/175036/RS// | sr |
dc.rights | openAccess | sr |
dc.rights.uri | https://creativecommons.org/licenses/by/4.0/ | |
dc.source | Biochemia Medica | |
dc.subject | quality control | sr |
dc.subject | moving average | sr |
dc.subject | bias detection simulation | sr |
dc.subject | MA Generator software | sr |
dc.title | Optimizing moving average control procedures for small-volume laboratories: can it be done? | sr |
dc.type | article | sr |
dc.rights.license | BY | sr |
dcterms.abstract | Игњатовић, Светлана; Лукић, Вера; Оптимизинг мовинг авераге цонтрол процедурес фор смалл-волуме лабораториес: цан ит бе доне?; Оптимизинг мовинг авераге цонтрол процедурес фор смалл-волуме лабораториес: цан ит бе доне?; | |
dc.citation.volume | 29 | |
dc.citation.issue | 3 | |
dc.citation.spage | 1 | |
dc.citation.epage | 13 | |
dc.citation.rank | M21 | |
dc.identifier.wos | 000489300500014 | |
dc.identifier.doi | 10.11613/BM.2019.030710 | |
dc.identifier.scopus | 2-s2.0-85073635870 | |
dc.identifier.fulltext | https://farfar.pharmacy.bg.ac.rs/bitstream/id/7272/Optimizing_moving_average_pub_2019.pdf | |
dc.type.version | publishedVersion | sr |