Sensitivity of three thyrotropin receptor antibody assays in thyroid-associated orbitopathy

Sarić-Matutinović, Marija; Diana, Tanja; Nedeljković-Beleslin, Biljana; Ćirić, Jasmina; Žarković, Miloš; Perović Blagojević, Iva; Kahaly, George; Ignjatović, Svetlana

(Society of Medical Biochemists of Serbia, 2022)

TY  - JOUR
AU  - Sarić-Matutinović, Marija
AU  - Diana, Tanja
AU  - Nedeljković-Beleslin, Biljana
AU  - Ćirić, Jasmina
AU  - Žarković, Miloš
AU  - Perović Blagojević, Iva
AU  - Kahaly, George
AU  - Ignjatović, Svetlana
PY  - 2022
UR  - https://farfar.pharmacy.bg.ac.rs/handle/123456789/4091
AB  - Background: Thyrotropin receptor autoantibodies (TSH-R-Ab) are indispensable biomarkers in the laboratory assessment of thyroid-associated orbitopathy (TAO). Clinical sensitivity of three different assays for TSH-R-Ab determination was evaluated in patients with TAO.

Methods: 87 consecutive TAO patients were enrolled and their serum samples analyzed in parallel with three assays. An ECLIA competitive binding and a chemiluminescent bridge immunoassay were used to measure total and bind-ing TSH-R-Ab concentration, while their functional activity was determined using a stimulatory TSH-R-Ab (TSAb) cell-based bioassay.

Results: Compared to the two binding assays (ECLIA p<0.001, bridge p=0.003), the TSAb bioassay was more sensitive pertaining to the positive detection of TSH-R-Ab in TAO patients. No difference (p=0.057) was noted between the ECLIA and bridge assays regarding sensitivity rate. All patients with active and/or moderate-to-severe TAO tested positive in the TSAb bioassay (100% and 100%, respectively), while the positivity rates for bridge and ECLIA binding assays were 89.7% and 82.1% for active TAO, and 90.2% and 86.3% for severe TAO, respectively. Negative predictive values of the bioassay, bridge, and ECLIA assays were 100%, 75%, and 71%, respectively for active TAO, and 100%, 86%, and 71%, respectively for moderate-to-severe TAO. The superiority of the bioassay was most prominent in euthyroid (ET) TAO. Positivity rates of the TSAb bioassay, bridge and ECLIA binding assays were 89.6%, 75%, and 64.6%, respectively for inactive TAO; 86.1%, 69.4%, and 52.8%, respectively for mild TAO; 87.5%, 62.5%, and 12.5%, respectively for euthyroid TAO. The bridge assay correlated better with the ECLIA binding assay (r=0.893, p<0.001), compared to the bioassay (r=0.669, p<0.001).

Conclusions: In patients with TAO of various activity and severity, the TSAb bioassay demonstrates a superior clinical performance compared to both ECLIA and bridge binding assays.
AB  - Uvod: Autoantitela na receptore za tireostimulišući hormon (TSH-R-Ab) su nezamenljivi biomarkeri u laboratorijskoj proceni orbitopatije udru'ene sa štitnom žlezdom (TAO). U radu je procenjena klinička osetljivost tri različita testa za određivanje TSH-R-Ab kod pacijenata sa TAO. Metode: U studiju je ključeno 87 uzastopnih pacijenata sa TAO i njihovi uzorci seruma su analizirani paralelno sa tri testa. Za merenje ukupne i vezujuće koncentracije TSH-RAb korišćeni su ECLIA imunohemijski test kompetitivnog vezivanja i hemiluminiscentni imunohemijski "sendvič" test, dok je njihova funkcionalna aktivnost određena pomoću ćelijskog biološkog testa (bioeseja) za određivanje stimulatornih TSH-R-Ab (TSAb). Rezultati: U poređenju sa dva imunohemijska testa vezivanja (ECLIA P<0,001, "sendvič test" P=0,003), TSAb bioesej se pokazao najosetljivijim u pogledu pozitivne detekcije TSH-RAb kod TAO pacijenata. Nikakva razlika nije detektovana (P=0,057) između ECLIA i "sendvič" testa u pogledu stope osetljivosti. TSAb bioesej je bio pozitivan kod svih pacijenata sa aktivnom i/ili umerenom do teškom TAO (redom 100% i 100%), dok su stope pozitivnosti za sendvič i ECLIA imunohemijski test bile 89,7% i 82,1% za aktivnu TAO, i 90,2% i 86,3% za tešku TAO, redom. Negativne prediktivne vrednosti bioeseja, "sendvič" i ECLIA testova bile su redom 100%, 75% i 71%, za aktivnu TAO, odnosno 100%, 86% i 71%, redom za umerenu do tešku TAO. Superiornost biološke analize bila je najistaknutija kod eutiroidnog oblika (ET) TAO. Stope pozitivnosti TSAb bioeseja, "sendvič" i ECLIA testova vezivanja bile su redom 89,6%, 75% i 64,6%, za neaktivnu TAO; 86,1%, 69,4% i 52,8% za blagu TAO; 87,5%, 62,5% i 12,5% za eutiroidni TAO. "Sendvič" imunohemijski test je bio u boljoj korelaciji sa ECLIA testom kompetitivnog vezivanja (r=0,893, P<0,001), u poređenju sa bioesejom (r=0,669, P<0,001). Zaključak: Kod pacijenata sa TAO različite aktivnosti i težine TSAb bioesej pokazuje superiorne kliničke performanse u poređenju sa obe vrste imunohemijskih testova vezivanja (ECLIA i "sendvič test").
PB  - Society of Medical Biochemists of Serbia
T2  - Journal of Medical Biochemistry
T1  - Sensitivity of three thyrotropin receptor antibody assays in thyroid-associated orbitopathy
T1  - Osetljivost tri testa za određivanje antitela na receptor za tireostimulišući hormon kod pacijenata sa orbitopatijom udruženom sa štitnom žlezdom
VL  - 41
IS  - 2
SP  - 211
EP  - 220
DO  - 10.5937/jomb0-34718
ER  - 
@article{
author = "Sarić-Matutinović, Marija and Diana, Tanja and Nedeljković-Beleslin, Biljana and Ćirić, Jasmina and Žarković, Miloš and Perović Blagojević, Iva and Kahaly, George and Ignjatović, Svetlana",
year = "2022",
abstract = "Background: Thyrotropin receptor autoantibodies (TSH-R-Ab) are indispensable biomarkers in the laboratory assessment of thyroid-associated orbitopathy (TAO). Clinical sensitivity of three different assays for TSH-R-Ab determination was evaluated in patients with TAO.

Methods: 87 consecutive TAO patients were enrolled and their serum samples analyzed in parallel with three assays. An ECLIA competitive binding and a chemiluminescent bridge immunoassay were used to measure total and bind-ing TSH-R-Ab concentration, while their functional activity was determined using a stimulatory TSH-R-Ab (TSAb) cell-based bioassay.

Results: Compared to the two binding assays (ECLIA p<0.001, bridge p=0.003), the TSAb bioassay was more sensitive pertaining to the positive detection of TSH-R-Ab in TAO patients. No difference (p=0.057) was noted between the ECLIA and bridge assays regarding sensitivity rate. All patients with active and/or moderate-to-severe TAO tested positive in the TSAb bioassay (100% and 100%, respectively), while the positivity rates for bridge and ECLIA binding assays were 89.7% and 82.1% for active TAO, and 90.2% and 86.3% for severe TAO, respectively. Negative predictive values of the bioassay, bridge, and ECLIA assays were 100%, 75%, and 71%, respectively for active TAO, and 100%, 86%, and 71%, respectively for moderate-to-severe TAO. The superiority of the bioassay was most prominent in euthyroid (ET) TAO. Positivity rates of the TSAb bioassay, bridge and ECLIA binding assays were 89.6%, 75%, and 64.6%, respectively for inactive TAO; 86.1%, 69.4%, and 52.8%, respectively for mild TAO; 87.5%, 62.5%, and 12.5%, respectively for euthyroid TAO. The bridge assay correlated better with the ECLIA binding assay (r=0.893, p<0.001), compared to the bioassay (r=0.669, p<0.001).

Conclusions: In patients with TAO of various activity and severity, the TSAb bioassay demonstrates a superior clinical performance compared to both ECLIA and bridge binding assays., Uvod: Autoantitela na receptore za tireostimulišući hormon (TSH-R-Ab) su nezamenljivi biomarkeri u laboratorijskoj proceni orbitopatije udru'ene sa štitnom žlezdom (TAO). U radu je procenjena klinička osetljivost tri različita testa za određivanje TSH-R-Ab kod pacijenata sa TAO. Metode: U studiju je ključeno 87 uzastopnih pacijenata sa TAO i njihovi uzorci seruma su analizirani paralelno sa tri testa. Za merenje ukupne i vezujuće koncentracije TSH-RAb korišćeni su ECLIA imunohemijski test kompetitivnog vezivanja i hemiluminiscentni imunohemijski "sendvič" test, dok je njihova funkcionalna aktivnost određena pomoću ćelijskog biološkog testa (bioeseja) za određivanje stimulatornih TSH-R-Ab (TSAb). Rezultati: U poređenju sa dva imunohemijska testa vezivanja (ECLIA P<0,001, "sendvič test" P=0,003), TSAb bioesej se pokazao najosetljivijim u pogledu pozitivne detekcije TSH-RAb kod TAO pacijenata. Nikakva razlika nije detektovana (P=0,057) između ECLIA i "sendvič" testa u pogledu stope osetljivosti. TSAb bioesej je bio pozitivan kod svih pacijenata sa aktivnom i/ili umerenom do teškom TAO (redom 100% i 100%), dok su stope pozitivnosti za sendvič i ECLIA imunohemijski test bile 89,7% i 82,1% za aktivnu TAO, i 90,2% i 86,3% za tešku TAO, redom. Negativne prediktivne vrednosti bioeseja, "sendvič" i ECLIA testova bile su redom 100%, 75% i 71%, za aktivnu TAO, odnosno 100%, 86% i 71%, redom za umerenu do tešku TAO. Superiornost biološke analize bila je najistaknutija kod eutiroidnog oblika (ET) TAO. Stope pozitivnosti TSAb bioeseja, "sendvič" i ECLIA testova vezivanja bile su redom 89,6%, 75% i 64,6%, za neaktivnu TAO; 86,1%, 69,4% i 52,8% za blagu TAO; 87,5%, 62,5% i 12,5% za eutiroidni TAO. "Sendvič" imunohemijski test je bio u boljoj korelaciji sa ECLIA testom kompetitivnog vezivanja (r=0,893, P<0,001), u poređenju sa bioesejom (r=0,669, P<0,001). Zaključak: Kod pacijenata sa TAO različite aktivnosti i težine TSAb bioesej pokazuje superiorne kliničke performanse u poređenju sa obe vrste imunohemijskih testova vezivanja (ECLIA i "sendvič test").",
publisher = "Society of Medical Biochemists of Serbia",
journal = "Journal of Medical Biochemistry",
title = "Sensitivity of three thyrotropin receptor antibody assays in thyroid-associated orbitopathy, Osetljivost tri testa za određivanje antitela na receptor za tireostimulišući hormon kod pacijenata sa orbitopatijom udruženom sa štitnom žlezdom",
volume = "41",
number = "2",
pages = "211-220",
doi = "10.5937/jomb0-34718"
}
Sarić-Matutinović, M., Diana, T., Nedeljković-Beleslin, B., Ćirić, J., Žarković, M., Perović Blagojević, I., Kahaly, G.,& Ignjatović, S.. (2022). Sensitivity of three thyrotropin receptor antibody assays in thyroid-associated orbitopathy. in Journal of Medical Biochemistry
Society of Medical Biochemists of Serbia., 41(2), 211-220.
https://doi.org/10.5937/jomb0-34718
Sarić-Matutinović M, Diana T, Nedeljković-Beleslin B, Ćirić J, Žarković M, Perović Blagojević I, Kahaly G, Ignjatović S. Sensitivity of three thyrotropin receptor antibody assays in thyroid-associated orbitopathy. in Journal of Medical Biochemistry. 2022;41(2):211-220.
doi:10.5937/jomb0-34718 .
Sarić-Matutinović, Marija, Diana, Tanja, Nedeljković-Beleslin, Biljana, Ćirić, Jasmina, Žarković, Miloš, Perović Blagojević, Iva, Kahaly, George, Ignjatović, Svetlana, "Sensitivity of three thyrotropin receptor antibody assays in thyroid-associated orbitopathy" in Journal of Medical Biochemistry, 41, no. 2 (2022):211-220,
https://doi.org/10.5937/jomb0-34718 . .
3
3