JavaScript is disabled for your browser. Some features of this site may not work without it.

Assessment of Liquid Chromatographic Method Robustness by Use of Plackett-Burman Design

Samo za registrovane korisnike
2010
Autori
Masković, M.
Jančić-Stojanović, Biljana
Malenović, Anđelija
Ivanović, D.
Medenica, Mirjana
Članak u časopisu (Objavljena verzija)
Metapodaci
Prikaz svih podataka o dokumentu
Apstrakt
Robustness testing is an important part of method validation. Many ideas on robustness testing can be found in the literature, but the most useful approach is application of experimental design. In the work discussed in this paper, Plackett-Burman design was used for robustness testing of an LC method. Robustness was assessed for a previously validated method developed for chromatographic analysis of perindopril tert-butylamine and its impurities. Eleven factors (seven real and four dummies) in twelve experiments were examined. Robustness was assessed by graphical (half-normal probability plots and Pareto charts) and statistical (t-test) methods. For estimation of the standard error of effect (necessary for t-test estimation) calculations based on negligible effects and Dong's algorithm were used at the significance levels alpha = 0.05 and alpha = 0.01. In this way the effect of the factors was completely defined and, furthermore, nonsignificant intervals for significant variables were... calculated. Finally, on the basis of the worst-case situation, system-suitability tests were performed and acceptance limits for certain values were calculated.

Ključne reči:
Plackett-Burman design / robustness / system-suitability tests / perindopril tert-butylamine / impurities / liquid chromatography
Izvor:
Acta Chromatographica, 2010, 22, 2, 281-296
Izdavač:
  • Akademiai Kiado Zrt, Budapest
Projekti:
  • Formulisanje i Karakterizacija separacionih sistema za modelovanje retencionog ponašanja lekovitih supstancija uz hemometrijsku evaluaciju (RS-142077)

DOI: 10.1556/AChrom.22.2010.2.10

ISSN: 1233-2356

WoS: 000279636700010

Scopus: 2-s2.0-77952962987
[ Google Scholar ]
16
12
URI
http://farfar.pharmacy.bg.ac.rs/handle/123456789/1399
Kolekcije
  • Radovi istraživača / Researchers’ publications
Institucija
Pharmacy
TY  - JOUR
AU  - Masković, M.
AU  - Jančić-Stojanović, Biljana
AU  - Malenović, Anđelija
AU  - Ivanović, D.
AU  - Medenica, Mirjana
PY  - 2010
UR  - http://farfar.pharmacy.bg.ac.rs/handle/123456789/1399
AB  - Robustness testing is an important part of method validation. Many ideas on robustness testing can be found in the literature, but the most useful approach is application of experimental design. In the work discussed in this paper, Plackett-Burman design was used for robustness testing of an LC method. Robustness was assessed for a previously validated method developed for chromatographic analysis of perindopril tert-butylamine and its impurities. Eleven factors (seven real and four dummies) in twelve experiments were examined. Robustness was assessed by graphical (half-normal probability plots and Pareto charts) and statistical (t-test) methods. For estimation of the standard error of effect (necessary for t-test estimation) calculations based on negligible effects and Dong's algorithm were used at the significance levels alpha = 0.05 and alpha = 0.01. In this way the effect of the factors was completely defined and, furthermore, nonsignificant intervals for significant variables were calculated. Finally, on the basis of the worst-case situation, system-suitability tests were performed and acceptance limits for certain values were calculated.
PB  - Akademiai Kiado Zrt, Budapest
T2  - Acta Chromatographica
T1  - Assessment of Liquid Chromatographic Method Robustness by Use of Plackett-Burman Design
VL  - 22
IS  - 2
SP  - 281
EP  - 296
DO  - 10.1556/AChrom.22.2010.2.10
ER  - 
@article{
author = "Masković, M. and Jančić-Stojanović, Biljana and Malenović, Anđelija and Ivanović, D. and Medenica, Mirjana",
year = "2010",
url = "http://farfar.pharmacy.bg.ac.rs/handle/123456789/1399",
abstract = "Robustness testing is an important part of method validation. Many ideas on robustness testing can be found in the literature, but the most useful approach is application of experimental design. In the work discussed in this paper, Plackett-Burman design was used for robustness testing of an LC method. Robustness was assessed for a previously validated method developed for chromatographic analysis of perindopril tert-butylamine and its impurities. Eleven factors (seven real and four dummies) in twelve experiments were examined. Robustness was assessed by graphical (half-normal probability plots and Pareto charts) and statistical (t-test) methods. For estimation of the standard error of effect (necessary for t-test estimation) calculations based on negligible effects and Dong's algorithm were used at the significance levels alpha = 0.05 and alpha = 0.01. In this way the effect of the factors was completely defined and, furthermore, nonsignificant intervals for significant variables were calculated. Finally, on the basis of the worst-case situation, system-suitability tests were performed and acceptance limits for certain values were calculated.",
publisher = "Akademiai Kiado Zrt, Budapest",
journal = "Acta Chromatographica",
title = "Assessment of Liquid Chromatographic Method Robustness by Use of Plackett-Burman Design",
volume = "22",
number = "2",
pages = "281-296",
doi = "10.1556/AChrom.22.2010.2.10"
}
Masković M, Jančić-Stojanović B, Malenović A, Ivanović D, Medenica M. Assessment of Liquid Chromatographic Method Robustness by Use of Plackett-Burman Design. Acta Chromatographica. 2010;22(2):281-296
Masković, M., Jančić-Stojanović, B., Malenović, A., Ivanović, D.,& Medenica, M. (2010). Assessment of Liquid Chromatographic Method Robustness by Use of Plackett-Burman Design.
Acta ChromatographicaAkademiai Kiado Zrt, Budapest., 22(2), 281-296.
https://doi.org/10.1556/AChrom.22.2010.2.10
Masković M., Jančić-Stojanović Biljana, Malenović Anđelija, Ivanović D., Medenica Mirjana, "Assessment of Liquid Chromatographic Method Robustness by Use of Plackett-Burman Design" 22, no. 2 (2010):281-296,
https://doi.org/10.1556/AChrom.22.2010.2.10 .