This paper investigates the ability of seven chromatographic response functions to measure the quality of chromatograms obtained in hydrophilic interaction liquid chromatography (HILIC). First, the functions were tested on a set of simulated chromatograms and differences in their mathematical design were discussed. Second, the functions were evaluated on the experimentally obtained chromatograms in HILIC analysis of model mixture consisted of beta agonists and antagonists. The ranking of chromatograms obtained by different functions was significantly different, implying that the accuracy of the optimization procedure is strongly dependent on the function that was selected as an output. Investigation of potential drawbacks of each function was conducted and general recommendations concerning the use of chromatographic response functions in optimization strategies are proposed.
Кључне речи:
Beta agonists and antagonists / Chromatographic response function (CRF) / HILICИзвор:
Analytical Letters, 2013, 46, 8, 1198-1212Издавач:
- Taylor & Francis Inc, Philadelphia
Финансирање / пројекти:
- Моделовање различитих хроматографских система са хемометријским приступом у фармацеутској анализи (RS-172052)
Напомена:
- Peer-reviewed manuscript: http://farfar.pharmacy.bg.ac.rs/handle/123456789/3419
DOI: 10.1080/00032719.2012.755689
ISSN: 0003-2719
WoS: 000318281700002
Scopus: 2-s2.0-84877274218
Институција/група
PharmacyTY - JOUR AU - Rakić, Tijana AU - Jančić-Stojanović, Biljana AU - Jovanović, Marko AU - Malenović, Anđelija AU - Ivanović, Darko AU - Medenica, Mirjana PY - 2013 UR - https://farfar.pharmacy.bg.ac.rs/handle/123456789/2003 AB - This paper investigates the ability of seven chromatographic response functions to measure the quality of chromatograms obtained in hydrophilic interaction liquid chromatography (HILIC). First, the functions were tested on a set of simulated chromatograms and differences in their mathematical design were discussed. Second, the functions were evaluated on the experimentally obtained chromatograms in HILIC analysis of model mixture consisted of beta agonists and antagonists. The ranking of chromatograms obtained by different functions was significantly different, implying that the accuracy of the optimization procedure is strongly dependent on the function that was selected as an output. Investigation of potential drawbacks of each function was conducted and general recommendations concerning the use of chromatographic response functions in optimization strategies are proposed. PB - Taylor & Francis Inc, Philadelphia T2 - Analytical Letters T1 - Evaluation of Seven Chromatographic Response Functions on Simulated and Experimentally Obtained Chromatograms in Hydrophilic Interaction Liquid Chromatography System VL - 46 IS - 8 SP - 1198 EP - 1212 DO - 10.1080/00032719.2012.755689 ER -
@article{ author = "Rakić, Tijana and Jančić-Stojanović, Biljana and Jovanović, Marko and Malenović, Anđelija and Ivanović, Darko and Medenica, Mirjana", year = "2013", abstract = "This paper investigates the ability of seven chromatographic response functions to measure the quality of chromatograms obtained in hydrophilic interaction liquid chromatography (HILIC). First, the functions were tested on a set of simulated chromatograms and differences in their mathematical design were discussed. Second, the functions were evaluated on the experimentally obtained chromatograms in HILIC analysis of model mixture consisted of beta agonists and antagonists. The ranking of chromatograms obtained by different functions was significantly different, implying that the accuracy of the optimization procedure is strongly dependent on the function that was selected as an output. Investigation of potential drawbacks of each function was conducted and general recommendations concerning the use of chromatographic response functions in optimization strategies are proposed.", publisher = "Taylor & Francis Inc, Philadelphia", journal = "Analytical Letters", title = "Evaluation of Seven Chromatographic Response Functions on Simulated and Experimentally Obtained Chromatograms in Hydrophilic Interaction Liquid Chromatography System", volume = "46", number = "8", pages = "1198-1212", doi = "10.1080/00032719.2012.755689" }
Rakić, T., Jančić-Stojanović, B., Jovanović, M., Malenović, A., Ivanović, D.,& Medenica, M.. (2013). Evaluation of Seven Chromatographic Response Functions on Simulated and Experimentally Obtained Chromatograms in Hydrophilic Interaction Liquid Chromatography System. in Analytical Letters Taylor & Francis Inc, Philadelphia., 46(8), 1198-1212. https://doi.org/10.1080/00032719.2012.755689
Rakić T, Jančić-Stojanović B, Jovanović M, Malenović A, Ivanović D, Medenica M. Evaluation of Seven Chromatographic Response Functions on Simulated and Experimentally Obtained Chromatograms in Hydrophilic Interaction Liquid Chromatography System. in Analytical Letters. 2013;46(8):1198-1212. doi:10.1080/00032719.2012.755689 .
Rakić, Tijana, Jančić-Stojanović, Biljana, Jovanović, Marko, Malenović, Anđelija, Ivanović, Darko, Medenica, Mirjana, "Evaluation of Seven Chromatographic Response Functions on Simulated and Experimentally Obtained Chromatograms in Hydrophilic Interaction Liquid Chromatography System" in Analytical Letters, 46, no. 8 (2013):1198-1212, https://doi.org/10.1080/00032719.2012.755689 . .