Како заиста видимо невидљиво: анализа утицаја услова мерења и припреме узорка на резултате процене величине наночестица/нанокапи применом динамичког расипања светлости

How do we actually see the invisible: analysis of the impact of measuring conditions and sample preparation procedure on the size estimation of nanoparticles/nanodroplets through dynamic light scattering

Thumbnail
2022
bitstream_9737.pdf (562.3Kb)
Authors
Paunović, Marija
Anđelković, Nastasija
Contributors
Savić, Snežana
Nikolić, Ines
Conference object (Published version)
Metadata
Show full item record
Abstract
Увод: Многе технике се користе за процену величине/дистрибуције величина у циљу карактеризације нанолекoва, a динамичко расипање светлости (DLS) представља једну од најчешће коришћених. Циљ рада: Циљ овог рада био је показати како различити услови мерења и начини припреме истог узорка могу дати различите резултате процене величине нанокапи/наночестица применом DLS методе, те понудити оптималaн протокол мерења. Материјал и методе: Као тест узорци коришћени су наноемулзија и водена дисперзија полимерних наночестица, a високопречишћена вода и изотонични фосфатни пуфер (PBS, pH 7,4) као медијуми за разблажење. Узорци су пре мерења разблаживани у 5 волуменских односа. Мерење је урађено применом batch-mode DLS, коришћењем уређаја ZetasizerNano ZS90, према NIST-EUNCL-PCC1 процедури. Праћени су следећи параметри: Z-ave (intensity-based diameter), атенуација, одсечак корелационе функције, број детектованих фотона, pH. Резултати: Примећено је да постоје статистички значајне разлике у добијени...м вредностима процењеног дијаметра нанокапи/наночестица које су последица степена разблажења и типа коришћеног медијума. Z-ave у пречишћеној води значајно је већи у односу на PBS при истом разблажењу истог узорка. Закључак: Праћењем релевантних параметара инструмента, предност је дата PBS-у као дисперзанту, а оптималан ниво разблажења за наночестице био је 1:10 (Z-ave=59,16±0,46 nm), а за наноемулзију 1:100 (Z-ave=73,5±0,75 nm). Hиво разблажења се мора индивидуално испитати за сваки узорак и не може се арбитрарно одређивати.

Introduction: Many techniques are used to estimate the size/size distribution in order to characterize nanodrugs, and dynamic light scatering (DLS) is one of the most commonly used. The Aim: The goal of this research was to show how different measurement conditions and sample preparation methods can give different results in terms of size estimation using DLS, and to offer an optimal measurement protocol. Material and Methods: Nanoemulsion and aqueous dispersion of polymer nanoparticles were used as test samples, whereas highly purified water and phosphate buffer saline (PBS, pH 7.4) were used as dilution media. Prior to measurements, samples were diluted in 5 volume ratios. The measurement was performed using batch-mode DLS, according to NIST-EUNCL-PCC1 procedure. The following parameters were monitored: Z-ave, attenuation, intercept of the autocorrelation function, counting rate of scattered light, pH. Results: Statistically significant differences in the obtained Z-ave values of ...nanodroplets/nanoparticles were observed with regard to the degree of dilution and the type of medium used. Z-ave in purified water was significantly higher than in PBS at the same dilution of the same sample. Conclusion: By monitoring the relevant instrument parameters, preference was given to PBS as dispersant, and the optimal dilution level for nanoparticles was 1:10 (Z-ave=59.16±0.46 nm) and for nanoemulsion 1:100 (Z-ave=73.5±0.75 nm). The level of dilution is sample-dependent and should not be arbitrarily determined.

Keywords:
dynamic light scattering / nanoemulsions / nanoparticles
Source:
2022
[ Google Scholar ]
Handle
https://hdl.handle.net/21.15107/rcub_farfar_4156
URI
https://farfar.pharmacy.bg.ac.rs/handle/123456789/4156
Collections
  • XV Mini-kongres studenata Farmaceutskog fakulteta „U svetu nauke i mladih istraživača”
Institution/Community
Pharmacy
TY  - CONF
AU  - Paunović, Marija
AU  - Anđelković, Nastasija
PY  - 2022
UR  - https://farfar.pharmacy.bg.ac.rs/handle/123456789/4156
AB  - Увод: Многе технике се користе за процену величине/дистрибуције величина у циљу карактеризације нанолекoва, a динамичко расипање светлости (DLS) представља једну од најчешће коришћених.
Циљ рада: Циљ овог рада био је показати како различити услови мерења и начини припреме истог узорка могу дати различите резултате процене величине нанокапи/наночестица применом DLS методе, те понудити оптималaн протокол мерења.
Материјал и методе: Као тест узорци коришћени су наноемулзија и водена дисперзија полимерних наночестица, a високопречишћена вода и изотонични фосфатни пуфер (PBS, pH 7,4) као медијуми за разблажење. Узорци су пре мерења разблаживани у 5 волуменских односа. Мерење је урађено применом batch-mode DLS, коришћењем уређаја ZetasizerNano ZS90, према NIST-EUNCL-PCC1 процедури. Праћени су следећи параметри: Z-ave (intensity-based diameter), атенуација, одсечак корелационе функције, број детектованих фотона, pH.
Резултати: Примећено је да постоје статистички значајне разлике у добијеним вредностима процењеног дијаметра нанокапи/наночестица које су последица степена разблажења и типа коришћеног медијума. Z-ave у пречишћеној води значајно је већи у односу на PBS при истом разблажењу истог узорка.
Закључак: Праћењем релевантних параметара инструмента, предност је дата PBS-у као дисперзанту, а оптималан ниво разблажења за наночестице био је 1:10 (Z-ave=59,16±0,46 nm), а за наноемулзију 1:100 (Z-ave=73,5±0,75 nm). Hиво разблажења се мора индивидуално испитати за сваки узорак и не може се арбитрарно одређивати.
AB  - Introduction: Many techniques are used to estimate the size/size distribution in order to characterize nanodrugs, and dynamic light scatering (DLS) is one of the most commonly used.
The Aim: The goal of this research was to show how different measurement conditions and sample preparation methods can give different results in terms of size estimation using DLS, and to offer an optimal measurement protocol.
Material and Methods: Nanoemulsion and aqueous dispersion of polymer nanoparticles were used as test samples, whereas highly purified water and phosphate buffer saline (PBS, pH 7.4) were used as dilution media. Prior to measurements, samples were diluted in 5 volume ratios. The measurement was performed using batch-mode DLS, according to NIST-EUNCL-PCC1 procedure. The following parameters were monitored: Z-ave, attenuation, intercept of the autocorrelation function, counting rate of scattered light, pH.
Results: Statistically significant differences in the obtained Z-ave values of nanodroplets/nanoparticles were observed with regard to the degree of dilution and the type of medium used. Z-ave in purified water was significantly higher than in PBS at the same dilution of the same sample.
Conclusion: By monitoring the relevant instrument parameters, preference was given to PBS as dispersant, and the optimal dilution level for nanoparticles was 1:10 (Z-ave=59.16±0.46 nm) and for nanoemulsion 1:100 (Z-ave=73.5±0.75 nm). The level of dilution is sample-dependent and should not be arbitrarily determined.
T1  - Како заиста видимо невидљиво: анализа утицаја услова мерења и припреме узорка на резултате процене величине наночестица/нанокапи применом динамичког расипања светлости
T1  - How do we actually see the invisible: analysis of the impact of measuring conditions and sample preparation procedure on the size estimation of nanoparticles/nanodroplets through dynamic light scattering
UR  - https://hdl.handle.net/21.15107/rcub_farfar_4156
ER  - 
@conference{
author = "Paunović, Marija and Anđelković, Nastasija",
year = "2022",
abstract = "Увод: Многе технике се користе за процену величине/дистрибуције величина у циљу карактеризације нанолекoва, a динамичко расипање светлости (DLS) представља једну од најчешће коришћених.
Циљ рада: Циљ овог рада био је показати како различити услови мерења и начини припреме истог узорка могу дати различите резултате процене величине нанокапи/наночестица применом DLS методе, те понудити оптималaн протокол мерења.
Материјал и методе: Као тест узорци коришћени су наноемулзија и водена дисперзија полимерних наночестица, a високопречишћена вода и изотонични фосфатни пуфер (PBS, pH 7,4) као медијуми за разблажење. Узорци су пре мерења разблаживани у 5 волуменских односа. Мерење је урађено применом batch-mode DLS, коришћењем уређаја ZetasizerNano ZS90, према NIST-EUNCL-PCC1 процедури. Праћени су следећи параметри: Z-ave (intensity-based diameter), атенуација, одсечак корелационе функције, број детектованих фотона, pH.
Резултати: Примећено је да постоје статистички значајне разлике у добијеним вредностима процењеног дијаметра нанокапи/наночестица које су последица степена разблажења и типа коришћеног медијума. Z-ave у пречишћеној води значајно је већи у односу на PBS при истом разблажењу истог узорка.
Закључак: Праћењем релевантних параметара инструмента, предност је дата PBS-у као дисперзанту, а оптималан ниво разблажења за наночестице био је 1:10 (Z-ave=59,16±0,46 nm), а за наноемулзију 1:100 (Z-ave=73,5±0,75 nm). Hиво разблажења се мора индивидуално испитати за сваки узорак и не може се арбитрарно одређивати., Introduction: Many techniques are used to estimate the size/size distribution in order to characterize nanodrugs, and dynamic light scatering (DLS) is one of the most commonly used.
The Aim: The goal of this research was to show how different measurement conditions and sample preparation methods can give different results in terms of size estimation using DLS, and to offer an optimal measurement protocol.
Material and Methods: Nanoemulsion and aqueous dispersion of polymer nanoparticles were used as test samples, whereas highly purified water and phosphate buffer saline (PBS, pH 7.4) were used as dilution media. Prior to measurements, samples were diluted in 5 volume ratios. The measurement was performed using batch-mode DLS, according to NIST-EUNCL-PCC1 procedure. The following parameters were monitored: Z-ave, attenuation, intercept of the autocorrelation function, counting rate of scattered light, pH.
Results: Statistically significant differences in the obtained Z-ave values of nanodroplets/nanoparticles were observed with regard to the degree of dilution and the type of medium used. Z-ave in purified water was significantly higher than in PBS at the same dilution of the same sample.
Conclusion: By monitoring the relevant instrument parameters, preference was given to PBS as dispersant, and the optimal dilution level for nanoparticles was 1:10 (Z-ave=59.16±0.46 nm) and for nanoemulsion 1:100 (Z-ave=73.5±0.75 nm). The level of dilution is sample-dependent and should not be arbitrarily determined.",
title = "Како заиста видимо невидљиво: анализа утицаја услова мерења и припреме узорка на резултате процене величине наночестица/нанокапи применом динамичког расипања светлости, How do we actually see the invisible: analysis of the impact of measuring conditions and sample preparation procedure on the size estimation of nanoparticles/nanodroplets through dynamic light scattering",
url = "https://hdl.handle.net/21.15107/rcub_farfar_4156"
}
Paunović, M.,& Anđelković, N.. (2022). Како заиста видимо невидљиво: анализа утицаја услова мерења и припреме узорка на резултате процене величине наночестица/нанокапи применом динамичког расипања светлости. .
https://hdl.handle.net/21.15107/rcub_farfar_4156
Paunović M, Anđelković N. Како заиста видимо невидљиво: анализа утицаја услова мерења и припреме узорка на резултате процене величине наночестица/нанокапи применом динамичког расипања светлости. 2022;.
https://hdl.handle.net/21.15107/rcub_farfar_4156 .
Paunović, Marija, Anđelković, Nastasija, "Како заиста видимо невидљиво: анализа утицаја услова мерења и припреме узорка на резултате процене величине наночестица/нанокапи применом динамичког расипања светлости" (2022),
https://hdl.handle.net/21.15107/rcub_farfar_4156 .

DSpace software copyright © 2002-2015  DuraSpace
About FarFaR - Pharmacy Repository | Send Feedback

OpenAIRERCUB
 

 

All of DSpaceCommunitiesAuthorsTitlesSubjectsThis institutionAuthorsTitlesSubjects

Statistics

View Usage Statistics

DSpace software copyright © 2002-2015  DuraSpace
About FarFaR - Pharmacy Repository | Send Feedback

OpenAIRERCUB